Catalog > Aerosols > Nebulizers > NL9M Atomisor nebulizer

Aerosols

NL9M Atomisor nebulizer

Product Overview

In accordance with the new European Standard NF EN 13544-1 (with an ATOMISOR ABOX+ aerosols compressor).
CE 0459 medical device
IIa Class
(DDM 93/42)

References for order:
- NL9M
- NL9MP (for children)

More...

New nebulizer equipped with Selector M.

The selector M uses the surface pressure and liquid gravity to reduce by a factor 2 the drug loss from nebulizer.

The selector M enables to reduce drug dumping in the atmosphere by using droplet impact principle in nebulizer.

Inhaling phase:
- Aerosol is produced by air decompression,
- Smallest particles are selected,
- Patient inhaled air improve aerosol production: breath enhanced nebulizer,
- Biggest accumulated particles are recycled at selector M bottom.

Exhalation phase:
- Exhausted particles in nebulizer are speeded up in selector M,
- Biggest particles return directly to liquid stage,
- The light fraction only is rejected out of the nebulizer.

Approved by:
- ASTRA FRANCE for PULMICORT-SUSPENSION
- ROCHE GENENTECH pour PULMOZYME connected to the compressor ref. AL.

Technical Details

For broncho-pulmonary ATOMISOR AEROSOLS COMPRESSORS

The ATOMISOR NL9M nebulizer owns exceptional performances validated according to the new European Standard NF EN 13544.1.

Performances according the Standard NF EN 13544-1(on ABOX+ compressor)

- Maximum capacity: 6 ml
- Minimum capacity: 1 ml

With a volume of 2 ml:
- MMAD: 3.9 µm,
- Liquid flow: 0.12 ml/min.,
- inhalable mass: 0.50 ml NaF 1%

Equipments:

- Tube 6 x 9 - length 1 m,
- Multi-positions mask R (adult) or RP (pediatric),
- Mouth end-piece C27D.
- Flask for dosage.

1 - CONNECTION:

- on ATOMISOR compressors only,
- for the Tobramycine, nebulization solution: compressor ATOMISOR ABOX+ exclusively.

2 - DISINFECTION and STERILIZATION:

- in boiling water (15 min. minimum), or it can be autoclaved at 134 °C.

The accessories (tube, mask, mouth end-piece) will be disinfected in boiling water at home (15 min.).

In hospitals it is recommended to change them after each use. In any case it is necessary to comply with the instructions on the packings and instructions for use.